Digital Display:
Waveform readout, string speed, help statements and instructions, and computer host information.
Flow Simulation Speeds:
10 to 200 centimeters per second, bi-directional.
Speed Drift:
Crystal-locked to 20 parts per million (0.002%).
Accuracy: +/- 1% of stated speed.
Pulsatile Waveforms:
16 pre-programmed and optional customer specified.
|
Waveforms Included:
Adult common carotid, stenotic carotid, femoral, aortic.
Fetal middle cerebral artery, renal artery, umbilical artery.
Pediatric descending thoracic artery, patent ductus arteriosus.
Test waveforms:
Sinewaves with peak speeds of 100, 150, and 200 cm/second. Triangle waves with peak speeds of 100, 150, and 200 cm/second. Stepped ramp wave with stops at 0, 20, 40, 60, 80, and 100 cm/sec.
Waveform Resolution:
Each waveform simulation contains 1000 points of resolution, or speed adjustments, enabling extremely complex simulation.
|
Computer Interface:
Industry standard
RS-232 interface built-in for future
enhancements and remote control.
Very useful for automated quality
control in a manufac- turing
environment.
Fluids Used in Tank:
Plain tap water (velocity 1480
meters per second at 20
degrees C) or velocity corrected
water/glycol solution
giving 1540 meters per
second at 20 degrees C.
Phantom adjusts
itself for either fluid.
Physical Specifications:
120 Volts AC, 50 Watts.
Total weight in travel case:
22 lbs. (10Kg).
Travel case dimensions:
17" x 17" 10" (25 x 19.5 x 14 cm).
Tank dimensions:
10" x 12" x 9" deep
(25 cm x 30 cm x 23 cm ).
|